Omnigrad S TC62
Explosion-proof thermocouple thermometer

Product picture of thermocouple thermometer TC62 ©Endress+Hauser

Safe monitoring of process temperatures in challenging applications e. g. in the Oil & Gas industry

The robust thermometer is designed for use in demanding and safety relevant applications e.g. in the Chemical, Oil & Gas and Energy industry. The sensor is designed for use in a separate thermowell. An optional head transmitter with all common communication protocols makes the device ready to use with enhanced measurement accuracy and reliability compared to directly wired sensors. The modular design with a variety of options can be adapted to the measurement task.

  • Előnyök

    • Types of protection for use in hazardous locations: Intrinsic safety (Ex ia), flameproof (Ex d), non-sparking (Ex nA)

    • High degree of flexibility thanks to modular design with standard terminal heads as per DIN EN 50446 and customer-specific immersion...

  • Alkalmazási terület

    • Heavy duty applications

    • Oil & Gas processing industry

    • Measuring range: –40 to 1 100 °C (–40 to 2 012 °F)

    • Protection class: IP68

Tulajdonságok és jellemzők

Vissza

Details

  • Mérési elv

    Thermocouple

  • Characteristic / Application

    US style

    modular temperature assembly

    for heavy duty applications

    suitable for hazardous areas

    threaded process connection

    with neck

    to use with thermowell

  • Terminal head

    TA21H – aluminium – screw cover w. chain

    TA21H – stainless steel – screw cover w. chain

    TA30H – aluminium – w. or w/o. display

    TA30H – aluminium – one or two cable entries NPT1/2", NPT3/4", M20x1.5

  • Thermowell / protection tube

    to use with thermowell

  • Insert / probe

    replaceable (without process interruption)

  • Outer diameter protection tube / Insert

    Insert Diameter

    6,0 mm (0,24'')

  • Max. immersion length on request

    up to 30.000,00 mm (1.181,10'')

  • Material protection tube/ thermowell

    Not defined

  • Optional coating

    Not defined

  • Process connection

    male thread:

    NPT1/2"

  • Tip shape

    straight

  • Surface roughness Ra

    0,8 μm (31,5 μin.)

  • Operating temperature range

    Type K:

    -40 °C ...1.100 °C

    (-40 °F ...2.012 °F)

    Type J:

    -40 °C ...750 °C

    (-40 °F ...1.382 °F)

  • Max. process pressure (static)

    at 20 °C: 500 bar (7.252 psi)

  • Accuracy

    class 1 acc. to IEC 60584

  • Response time

    depending on configuration

    t50 = 3 s

    t90 = 7 s

  • Integration head transmitter

    yes (4 … 20 mA; HART; PROFIBUS PA; FOUNDATION FIELDBUS)

  • Ex - approvals

    ATEX II

    IECEx

    NEPSI

    EAC Ex

    Explosion proof

  • Certification

    Gost Metrology

    SIL (transmitter only)

Dokumentumok / Kézikönyvek / Szoftver

Tartozékok / alkatrészek

Nr
Order code
Specification
  • Előd