Próbáljon kulcsszavak, rendelési- vagy termékkód, vagy gyári szám szerint keresni, pl. „CM442” vagy „Műszaki információk”
A keresés elindításához írjon be legalább 2 karaktert.
  • Előzmények
The CA72TOC analyzer determines the total organic carbon (TOC) content in a number of industries.

High temperature TOC analyzer CA72TOC

Reliable online monitoring of industrial wastewater

CA72TOC
Please Wait (spinning wheel)
Ár feltöltés alatt
Az árak ideiglenesen nem érhetők el
Price on request
from CHF ??.-
Mérési feladat/Alkalmazás szűrése
Mérési feladat / Alkalmazás

Konfigurálja a termék CAD-rajzainak letöltéséhez

  • Mérési elv

    UV photometric

  • Characteristic

    Analyzer for online-determination of TOC in aqueous media using thermic catalytic combustion

  • Size

    1558 x 785 x 400 mm
    61.3 x 30.9 x 15.8 inch

  • Design

    IP54

  • Process temperature

    0 to 40 °C (32 to 104 °F)

  • Ambient temperature

    5 to 35 °C (41 to 95 °F)

  • Process pressure

    Pressure free

  • Sample flow rate

    20 ml/min (0.32 US gal/hr)

  • Consistency of the sample

    Aqueous
    Flammable substances are not allowed in ignitable concentrations, in such cases a sample dilution is required

  • Specials

    Heatable Salt Trap
    Exchange furnace concept
    pH-controlled acid dosage with TIC stripping

  • Application

    Industrial wastewater monitoring
    Control of process wastewater
    Monitoring of industrial surface water
    Recording of the carbon load to control the nutrient dosage

  • Power supply

    115 VAC, 230 VAC, 50/60Hz

  • Output / communication

    0/4 to 20 mA, galvanically separated
    Two channel version has an addtional one
    RS 232 C, propietary, for data output and remote control (optional)

  • Input

    8 signal inputs 24 V DC, ative, load max. 500 Ohm

  • Measurement range

    0.25 to 600 mg TOC/l
    1 to 2400 mg TOC/l
    2.5 to 6000 mg TOC/l
    5 to 12000 mg TOC/l
    With the optional pre-dilution the measuring range will expand by the factor of 20

  • Consumables

    Stripp reagent, KHP or citric acid standard