Aluminum analyzer
Liquiline System CA80AL

Liquiline System CA80AL - High-precision process analyzer for aluminum measurement ©Endress+Hauser
Liquiline System CA80AL aluminum analyzer - open view ©Endress+Hauser
Sample collector with level detection for the monitoring of sample transport ©Endress+Hauser

Aluminum concentration in a liquid can be measured using the colorimetric measuring principle. This video shows what it is about and how this measuring principle works.

Colorimetric system for drinking water and wastewater monitoring and optimization of process control

Liquiline System CA80AL offers precise, regulation-compliant online measurement of aluminum residues in drinking water and wastewater. The analyzer helps you to optimize aluminum dosing in flocculation processes and phosphate removal. It enables plug & play of up to 4 Memosens sensors – minimizing the installation effort. Automatic calibration and cleaning and the low reagent consumption reduce operating costs while advanced diagnostics with remote access ensure process safety.

  • Előnyök

    • The aluminum analyzer uses the pyrocatechol violet method according to DIN ISO 10566, ensuring direct comparability to lab results

    • Supports predictive maintenance by advanced diagnostics

    • Reduces operating costs thanks to automatic calibration and...

  • Alkalmazási terület

    Liquiline System CA80AL aluminum analyzer supports:

    • Limit monitoring in drinking water

    • Optimization of aluminum-based coagulation processes

    • Phosphate elimination in wastewater treatment plants

Tulajdonságok és jellemzők

Vissza

Details

  • Mérési elv

    Colorimetric

  • Characteristic

    Process analyzer for aluminum in aqueous solutions

    Comply with standard colorimetric measuring principle - pyrocatechol violet methot

  • Size

    Cabinet
    793 x 530 x 417 mm
    31.22 x 20.87 x 16.42 in
    Stand housing
    1693

  • Design

    Open design, cabinet and stand housing
    High-Performance plastic ASA-PC, additional stand coated steel

  • Process temperature

    4 ... 40 °C (39.2 ... 104 °F)

  • Ambient temperature

    5 ... 40C ( 41 ... 104F)

  • Process pressure

    at atmospheric pressure, < 0.2 bar

  • Sample flow rate

    min. 5ml/min (0,17 fl.oz/min)

  • Consistency of the sample

    low solid content (< 50NTU), aqueous homogeneous

  • Application

    Monitoring of the aluminum content of potable water, mineral water and industrial waste water

  • Power supply

    100 ... 120 VAC / 200 ... 240 VAC ± 10%
    24 VDC ± 10%
    50 ± 1 oder 60 ± 1,2 Hz

  • Output

    2x 0/4 ... 20 mA
    Webserver, Modbus, Ethernet/IP, Profibus DP

  • Input

    1 or 2 measuring channel

    1 ... 4 digital sensor inputs for sensors with Memosens protocol (optional)

  • Measurement range

    15 ... 1000 µg/l

  • Consumables

    Reagents and standard solutions CY80AL as well as cleaner CY800 are necessary for the operation. Regular maintenance is done with the parts of the maintenance kit CAV800.

Dokumentumok / Kézikönyvek / Szoftver

Tartozékok / alkatrészek

Nr
Order code
Specification